Name:

MST-KT300 300mA Medical X-ray Camera

 

Name:

MST-KT300 300mA Medical X-ray Camera

Code:MST1073356
PRODUCTS INTRODUCTION:

USE:General Photograph Ray-filter Photograph

MAIN CHARACTERISTICS :

1.Single bed and single bulb.

2.Design according to the electric mechanism of middle-size machine.The function is full and reasonable.

3.The high-tension system is of march-past structure.Equipped with afullwave rectification high-voltage generator and independent X-ray tube unit,with large output power and reliable and stable operation.

4.Supply voltage V,roentgenoscopy,photograpy KV and stepless adjustable.

5.Equipped with the manostat for filament of X-ray tube and the spacecharge compensator as to ensure the stable and accurate output.

6.Equipped with the capacity limitation and protection circuit and KvmA sinterlock limitation and protection.

7.Adopt the digital circuit timer,grade according to R10 prioritycoefficient and can control time exactly

8.High-voltage primary uses the zero controlled circuit of silicon-controlled retctifier of large power.

9.Photographic bed can move in length and breadth

10.The photographic bed,pillar and vibarating ray-filter are in a whole without top and bottom track.

MAIN TECHNICAL INDEXES
ITEM CONTENT INDEX
Supply power

capacity

Two-phase 30KVA

voltage frequency 380V/220V ± 10% 50Hz ± 0.5Hz

inner resistance

0.75 Ω at 380V, 0.25 Ω at 220V

Power supply

Adjustable range

380V,380V ± 10% ,220V 时 ,220V ± 10%,being adjustable continuously

Photograpy

voltage

50~100KV* 、 50~125KV Be adjustable continuously

current

small focus 50mA 100mA and big focus 50 、 100 、 200 、 300mA

time

0.04~6.3s,23 grades together arading according to R10 coefficient

High-Voltage generator

capacity

20KVA(instantaneous)

Max. Dc output voltage 100KV *125KV
Max. Dc output current 300mA

X-Ray tube

model

XD51-2040/100* XD51-2040/125

focus

small focus 1X1mm big focus 2X2mm

hotographic bed(YZ-I)

Bed face(LXWXH)

2000mmX665X700mm

Moving range of bed face

640mm in length; 200mm in breadth

Ray-filter of bed

Travel in length >=500mm

rid density N28

Convergence distance fo 100cm

Grid ratio R8

Pillar of X-ray source unit

Moving in length along photographic bed

1800mm or 1600mm

Distance around as moving up and down(along upright post)

650-1800mm

Rotary around the center of cross arm

± 180°

Rotary around the shaft line of X-ray annular tubes

-10 ° ~ 60° ~ +120°
Max size of cassette for photograph 356mmX432mm(14”X17”)